Use of medium cut-off membrane dialyzer compared to high flux membrane dialyzer in improving the clearance of middle to large molecule uremic toxins: A Systematic Review

DOI: https://doi.org/10.33024/minh.v8i3.934

Authors

  • Wasal Desrial Siregar Universitas Indonesia
  • Krisna Yetti universitas indonesia
  • Agung Waluyo universitas indonesia
  • yulia yulia universitas indonesia

High Flux Large Middle Molecules Medium Cut-Off Membrane Uremic Toxins

Abstract

Background: The loss of kidney function in patients with kidney disease leads to the accumulation of dissolved substances known as uremic toxins, which can negatively impact patient health. Membranes with a high number of pores, such as those in high-flux dialyzers, allow some medium-sized molecules like beta-2 microglobulin to pass through. However, these membranes are less effective at filtering out larger dissolved substances. Larger medium-sized molecules (>25 kD) need to be removed either by convection or through the use of highly permeable membranes such as Medium Cut-Off (MCO) membranes. Therefore, it is important to understand the improvement in uremic toxin clearance when using MCO dialyzers compared to high-flux dialyzers.

Purpose: To evaluate and compare the effectiveness of using medium cut-off (MCO) membrane dialyzer with high flux membrane dialyzer.

Method: A systematic review with article selection guided by the Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) guidelines. The systematic review process begins with the formulation of clinical questions relevant to the topic. Before this, the authors establish PICOS criteria, which stand for: P (problem, patient, or population), I (intervention, prognostic factor, or exposure), C (comparison or control), O (outcome), and S (study design). For this article, P: Patients with chronic kidney disease undergoing hemodialysis, I: Use of dialyzers with medium cut-off (MCO) membrane, C: Use of dialyzer with high flux membrane, O: Increased clearance of uremic toxins of middle to large molecules, such as β2-microglobulin, kappa and lambda free light chains, S: Randomized controlled trials. The article search was conducted systematically using the keywords: "Medium cut-off dialyzer" AND "High flux dialyzer" AND "Uremic toxins" AND "Middle molecules" AND "Large molecules" AND "Clearance Hemodialysis".

Results: The use of MCO membranes was more effective in eliminating medium-to-large uremic toxins while maintaining safety and regulation comparable to conventional hemodialysis with high-flux dialyzers. MCO membranes were also more effective even with low blood flow, without significantly increasing albumin loss. Additionally, MCO membranes contributed to reducing inflammation and allowed for shorter recovery times post-hemodialysis treatment.

Conclusion: Given its comparable safety to high-flux dialyzers, the use of MCO membranes can be expanded in clinical practice at hemodialysis centers, including those in resource-limited settings.

Downloads

Download data is not yet available.

References

Aleksova, J., Milat, F., Kotowicz, M. A., Pasco, J. A., Schultz, C., Wong, P., & Elder, G. J. (2021). Patients With End-Stage Kidney Disease Have Markedly Abnormal Cortical Hip Parameters By Dual-Energy X-Ray Absorptiometry. Nephrology Dialysis Transplantation, 36(3), 543-550.

Belmouaz, M., Bauwens, M., Hauet, T., Bossard, V., Jamet, P., Joly, F., & Bridoux, F. (2020). Comparison Of The Removal Of Uraemic Toxins With Medium Cut-Off And High-Flux Dialysers: A Randomized Clinical Trial. Nephrology Dialysis Transplantation, 35(2), 328-335.

Bolton, S., Gair, R., Nilsson, L. G., Matthews, M., Stewart, L., & Mccullagh, N. (2021). Clinical Assessment Of Dialysis Recovery Time And Symptom Burden: Impact Of Switching Hemodialysis Therapy Mode. Patient Related Outcome Measures, 315-321.

Boschetti-De-Fierro, A., Voigt, M., Storr, M., & Krause, B. (2013). Extended Characterization Of A New Class Of Membranes For Blood Purification: The High Cut-Off Membranes. The International Journal Of Artificial Organs, 36(7), 455-463.

Boschetti-De-Fierro, A., Voigt, M., Storr, M., & Krause, B. (2015). MCO Membranes: Enhanced Selectivity In High-Flux Class. Scientific Reports, 5(1), 18448.

Cobob, M. J. G., & Lindholm, B. (2017). Wasting In The Uremic Milieu. Expanded Hemodialysis: Innovative Clinical Approach In Dialysis, 191, 58-71.

Eiamcharoenying, J., Takkavatakarn, K., Chariyavilaskul, P., Susantitaphong, P., Eiam-Ong, S., & Tiranathanagul, K. (2022). Comparative Effectiveness Between Expanded Hemodialysis (Hemodialysis Using A Medium Cut-Off Dialyzer) And Mixed-Dilution Online Hemodiafiltration Using A High-Flux Dialyzer In Removing Middle-Molecule Uremic Toxins. Blood Purification, 51(11), 923-931.

Fiorentino, M., La Fergola, F., & De Rosa, S. (2024). Medium Cut-Off Dialyzer For Middle Molecular Uremic Toxins In AKI And Chronic Dialysis. Journal Of Nephrology, 37(1), 23-37.

García-Prieto, A., Vega, A., Linares, T., Abad, S., Macías, N., Aragoncillo, I., & Luno, J. (2018). Evaluation Of The Efficacy Of A Medium Cut-Off Dialyser And Comparison With Other High-Flux Dialysers In Conventional Haemodialysis And Online Haemodiafiltration. Clinical Kidney Journal, 11(5), 742-746.

Kandi, M., Brignardello-Petersen, R., Couban, R., Wu, C., & Nesrallah, G. (2022). Clinical Outcomes With Medium Cut-Off Versus High-Flux Hemodialysis Membranes: A Systematic Review And Meta-Analysis. Canadian Journal Of Kidney Health And Disease, 9, 20543581211067087.

Kim, T. H., Kim, S. H., Kim, T. Y., Park, H. Y., Jung, K. S., Lee, M. H., & Park, H. C. (2020). Removal Of Large Middle Molecules Via Haemodialysis With Medium Cut-Off Membranes At Lower Blood Flow Rates: An Observational Prospective Study. BMC Nephrology, 21, 1-9.

Mjøen, G., Reisaeter, A., Hallan, S., Line, P. D., Hartmann, A., Midtvedt, K., & Holdaas, H. (2016). Overall And Cardiovascular Mortality In Norwegian Kidney Donors Compared To The Background Population. Nephrology Dialysis Transplantation, 31(1), 170-170.

Nicholas, S. B., Alicic, R. Z., & Shen, J. (2022). Disparities In The Use Of SGLT2 Inhibitors And Glucagon-Like Peptide-1 Receptor Agonists In Adults With CKD In The United States. Kidney Medicine, 5(1), 100590.

Santos Jr, A. H., Chen, C., Casey, M. J., Womer, K. L., & Wen, X. (2018). New-Onset Diabetes After Kidney Transplantation: Can The Risk Be Modified By Choosing Immunosuppression Regimen Based On Pretransplant Viral Serology?. Nephrology Dialysis Transplantation, 33(1), 177-184.

Sevinc, M., Hasbal, N. B., Yilmaz, V., Basturk, T., Ahbap, E., Sakaci, T., & Unsal, A. (2020). Comparison Of Circulating Levels Of Uremic Toxins In Hemodialysis Patients Treated With Medium Cut-Off Membranes And High-Flux Membranes: Theranova In Sisli Hamidiye Etfal (THE SHE) Randomized Control Study. Blood Purification, 49(6), 733-742.

Van De Perre, E., Jones, R. B., & Jayne, D. R. (2021). Iga Vasculitis (Henoch–Schönlein Purpura): Refractory And Relapsing Disease Course In The Adult Population. Clinical Kidney Journal, 14(8), 1953-1960.

Weiner, D. E., Falzon, L., Skoufos, L., Bernardo, A., Beck, W., Xiao, M., & Tran, H. (2020). Efficacy And Safety Of Expanded Hemodialysis With The Theranova 400 Dialyzer: A Randomized Controlled Trial. Clinical Journal Of The American Society Of Nephrology, 15(9), 1310-1319.

Wolley, M., Jardine, M., & Hutchison, C. A. (2018). Exploring The Clinical Relevance Of Providing Increased Removal Of Large Middle Molecules. Clinical Journal Of The American Society Of Nephrology, 13(5), 805-814.

Yamamoto, S., Kazama, J. J., Wakamatsu, T., Takahashi, Y., Kaneko, Y., Goto, S., & Narita, I. (2016). Removal Of Uremic Toxins By Renal Replacement Therapies: A Review Of Current Progress And Future Perspectives. Renal Replacement Therapy, 2, 1-8.

Yang, J., Ke, G., Liao, Y., Guo, Y., & Gao, X. (2022). Efficacy Of Medium Cut‐Off Dialyzers And Comparison With High‐Flux Dialyzers In Patients On Maintenance Hemodialysis: A Systematic Review And Meta‐Analysis. Therapeutic Apheresis And Dialysis, 26(4), 756-768.

Zhou, X. L., Chen, Y. H., & Wang, Q. Y. (2017). A New Approach Combining Venoarterial Extracorporeal Membrane Oxygenation And CRRT For Adults: A Retrospective Study. The International Journal Of Artificial Organs, 40(7), 345-349.

Zickler, D., Schindler, R., Willy, K., Martus, P., Pawlak, M., Storr, M., & Girndt, M. (2017). Medium Cut-Off (MCO) Membranes Reduce Inflammation In Chronic Dialysis Patients—A Randomized Controlled Clinical Trial. Plos One, 12(1), E0169024.

Downloads

Published

2025-05-29

How to Cite

Use of medium cut-off membrane dialyzer compared to high flux membrane dialyzer in improving the clearance of middle to large molecule uremic toxins: A Systematic Review. (2025). MIN, 8(3), 324-332. https://doi.org/10.33024/minh.v8i3.934